1 Method Overview
After microwave digestion, the sample solution is injected into the graphite furnace of the atomic absorption spectrophotometer. After electrothermal atomization, it absorbs the resonance line at 283.3 nm. Within a certain concentration range, the absorbance value is proportional to the lead content, and quantification is performed by comparison with the standard series.
2 Scope of Application
This method is applicable to the determination of lead in olive oil.
3 Reference Basis
GB 5009.12‑2017
4 Instruments and Reagents
4.1 Instruments and Equipment
4.1.1 Testing Instruments
| No. | Name | Quantity | Technical Requirement | Accessories |
|---|---|---|---|---|
| 1 | Graphite Furnace Atomic Absorption Spectrophotometer | 1 set | — | Lead Hollow‑Cathode Lamp |
| 2 | Argon | 1 cylinder | Purity ≥99.99% | — |
| 3 | Circulating Cooling Water System | 1 set | — | — |