Determination of Iron in Graphite Powder (Flame Atomic Absorption Spectrometry)

1 Method Overview After the sample is digested with aqua regia, the measurement is performed using an air‑acetylene flame at the instrument wavelength of 248.3 nm.

2 Scope of Application This method is applicable to the determination of iron in graphite powder.

3 Reference Basis GB/T 24533‑2009

4 Instruments and Reagents

4.1 Instruments and Equipment

4.1.1 Testing Instruments

No. Name Quantity Technical Requirement Accessories
1 Flame Atomic Absorption Spectrophotometer 1 set Iron Hollow‑Cathode Lamp
2 Acetylene 1 cylinder Purity ≥99.99%
3 Air Compressor 1 set Rated exhaust pressure: 0.3 MPa