1 Method Overview

The wheel hub is thoroughly cleaned with ultrapure water and air-dried. A scraper is used to remove the topcoat, intermediate coating and primer on the surface. After microwave digestion, the concentration of chromium is determined by flame atomic absorption spectrometry at a wavelength of 357.9 nm.

2 Scope of Application

Determination of chromium in surface coatings of wheel hubs

3 Reference Standard

IEC 62312

4 Instruments and Reagents

4.1 Instruments and Equipment

4.1.1 Analytical Instruments

表格

No. Name Quantity Technical Requirements Accessories
1 Flame Atomic Absorption Spectrophotometer 1 set Chromium Hollow Cathode Lamp
2 Acetylene Gas Cylinder 1 cylinder Purity ≥ 99.99%
3 Air Compressor 1 unit Rated exhaust pressure: 0.3 MPa