1 Yöntem Genel Bakış

The wheel hub is thoroughly cleaned with ultrapure water and air-dried. A scraper is used to remove the topcoat, intermediate coating and primer on the surface. After microwave digestion, the concentration of chromium is determined by flame atomic absorption spectrometry at a wavelength of 357.9 nm.

2. Uygulama Kapsamı

Determination of chromium in surface coatings of wheel hubs

3 Referans Standardı

IEC 62312

4 Instruments and Reagents

4.1 Aletler ve Ekipmanlar

4.1.1 Analytical Instruments

表格

- Hayır. - Hayır. İsim Miktar Teknik Gereksinimler Aksesuarlar
1 Alev Atom Emme Spektrofotometri 1 set Krom Boş Katod Lambası
2 Acetylene Gas Cylinder 1 silindir Saflık ≥% 99,99
3 Hava Kompresörü 1 birim Nominal egzoz basıncı: 0.3 MPa