1 Panoramica del metodo

The wheel hub is thoroughly cleaned with ultrapure water and air-dried. A scraper is used to remove the topcoat, intermediate coating and primer on the surface. After microwave digestion, the concentration of chromium is determined by flame atomic absorption spectrometry at a wavelength of 357.9 nm.

2 Ambito di applicazione

Determination of chromium in surface coatings of wheel hubs

3 Norma di riferimento

IEC 62312

4 Instruments and Reagents

4.1 Strumenti e attrezzature

4.1.1 Analytical Instruments

表格

- No, no, no. Nome Quantità Requisiti tecnici Accessori
1 Spettrofotometro di assorbimento atomico di fiamma 1 set Lampada a catodo cavo al cromo
2 Acetylene Gas Cylinder 1 cilindro Purezza ≥ 99,99%
3 compressore aria 1 unità Pressione nominale di scarico: 0,3 MPa