1 Aperçu de la méthode

The wheel hub is thoroughly cleaned with ultrapure water and air-dried. A scraper is used to remove the topcoat, intermediate coating and primer on the surface. After microwave digestion, the concentration of chromium is determined by flame atomic absorption spectrometry at a wavelength of 357.9 nm.

2. Champ d'application

Determination of chromium in surface coatings of wheel hubs

3 Norme de référence

IEC 62312

4 Instruments and Reagents

4.1 Instruments et équipements

4.1.1 Analytical Instruments

表格

- Non, non. Nom Quantité Exigences techniques Accessoires
1 Spectrophotomètre d'absorption atomique de flamme 1 ensemble Lampe à cathode creuse au chrome
2 Acetylene Gas Cylinder 1 cylindre Pureté ≥ 99,99%
3 Compresseur d'air 1 unité Pression nominale d'échappement: 0,3 MPa