올리브 오일 (Graphite Oven Atomic Absorption Spectrometry)

1 방법 개요
After microwave digestion, the sample solution is injected into the graphite furnace of the atomic absorption spectrophotometer. After electrothermal atomization, it absorbs the resonance line at 283.3 nm. Within a certain concentration range, the absorbance value is proportional to the lead content, and quantification is performed by comparison with the standard series.

2. 적용 범위
This method is applicable to the determination of lead in olive oil.

3 Reference Basis
GB 5009.12‑2017

4. 기기 및 시약

4.1 기기 및 장비

4.1.1 Testing Instruments

아니야 이름 수량 기술 요구사항 부속품
1 그래파이트 로브 원자 흡수 분광광계는 1 세트 Lead Hollow‑Cathode Lamp
2 Argon 1 실린더 Purity ≥99.99%
3 순환 냉각물 시스템 1 세트