1 Yöntem Genel Bakış
After microwave digestion, the sample solution is injected into the graphite furnace of the atomic absorption spectrophotometer. After electrothermal atomization, it absorbs the resonance line at 283.3 nm. Within a certain concentration range, the absorbance value is proportional to the lead content, and quantification is performed by comparison with the standard series.
2. Uygulama Kapsamı
This method is applicable to the determination of lead in olive oil.
3 Reference Basis
GB 5009.12‑2017
4 Alet ve Reaktifler
4.1 Aletler ve Ekipmanlar
4.1.1 Testing Instruments
| - Hayır. - Hayır. | İsim | Miktar | Teknik Gereksinim | Aksesuarlar |
|---|---|---|---|---|
| 1 | Grafit Fırın Atom Emme Spektrofotometri | 1 set | — | Lead Hollow‑Cathode Lamp |
| 2 | Argon | 1 silindir | Purity ≥99.99% | — |
| 3 | Sirkülasyon Soğutma Su Sistemi | 1 set | — | — |