Süt Tozu'ndaki Kalsiyumun Belirlenmesi (Alev Atom Emme Spektrometrisi)

1 Yöntem Genel Bakış
After the sample is digested, lanthanum solution is added as a releasing agent. The sample is atomized in an atomic‑absorption flame, and the absorbance is measured at 422.7 nm. Within a certain concentration range, the absorbance is proportional to the calcium content, and quantification is performed by comparison with the standard series.

2. Uygulama Kapsamı
This method is applicable to the determination of calcium in milk powder.

3 Reference Basis
GB 5009.92‑2016, National Food Safety Standard — Determination of Calcium in Foods

4 Alet ve Reaktifler

4.1 Aletler ve Ekipmanlar

4.1.1 Testing Instruments

- Hayır. - Hayır. İsim Miktar Teknik Gereksinim Aksesuarlar
1 Alev Atom Emme Spektrofotometri 1 set Calcium Hollow‑Cathode Lamp
2 Hava Kompresörü 1 set Rated exhaust pressure: 0.3 MPa
3 Asetilen 1 silindir Purity ≥99.99%