1 Обзор метода
After the sample is digested, lanthanum solution is added as a releasing agent. The sample is atomized in an atomic‑absorption flame, and the absorbance is measured at 422.7 nm. Within a certain concentration range, the absorbance is proportional to the calcium content, and quantification is performed by comparison with the standard series.
2. Область применения
This method is applicable to the determination of calcium in milk powder.
3 Reference Basis
GB 5009.92‑2016, National Food Safety Standard — Determination of Calcium in Foods
4. Приборы и реагенты
4.1 Инструменты и оборудование
4.1.1 Testing Instruments
| Нет, нет. | Имя | Количество | Технические требования | Аксессуары |
|---|---|---|---|---|
| 1 | Спектрофотометр атомной абсорбции пламени | 1 набор | — | Calcium Hollow‑Cathode Lamp |
| 2 | Воздушный компрессор | 1 набор | Rated exhaust pressure: 0.3 MPa | — |
| 3 | ацетилен | 1 цилиндр | Purity ≥99.99% | — |