1 Descripción general del método
After pretreatment, the sample solution is introduced into the atomic absorption spectrophotometer for flame atomization, and the absorbance is measured. Within a certain concentration range, absorbance is proportional to nickel content, and quantification is performed by comparison with a standard series.
2 Ámbito de aplicación
This method applies to the determination of nickel in the reflective coating of mirrors.
3 Reference Standards
Ninguno
4 Instruments, Equipment and Reagents
4.1 Instrumentos y equipos
4.1.1 Analytical Instruments
表格
| ¡No! No! | Nombre | Cantidad | Requisitos técnicos | Accesorios |
|---|---|---|---|---|
| 1 | Espectrofotómetro de absorción atómica de llama | 1 conjunto | — | Lámpara catódica hueca de níquel |
| 2 | Compresor de aire | 1 unidad | Presión nominal de escape: 0,3 MPa | — |
| 3 | Acetylene Cylinder | 1 cilindro | Pureza ≥ 99,99% | — |