1 Descripción general del método
After sample digestion, the sample is atomized by flame, and the absorbance is measured at 324.8 nm. Within a certain concentration range, the absorbance of copper is proportional to its content, and quantification is performed by comparison with the standard series.
2 Ámbito de aplicación
This method is applicable to the determination of copper in nylon.
3 Instruments and Reagents
3.1 Instrumentos y equipos
3.1.1 Testing Instruments
| ¡No! No! | Nombre | Cantidad | Requisito técnico | Accesorios |
| 1 | Espectrofotómetro de absorción atómica de llama | 1 conjunto | — | Copper Hollow-Cathode Lamp |
| 2 | Compresor de aire | 1 conjunto | Presión nominal de escape: 0,3 MPa | — |
| 3 | Acetileno | 1 cilindro | Purity ≥99.99% | — |