Determinazione del rame in nylon (spettrometria di assorbimento atomico di fiamma)

1 Panoramica del metodo

After sample digestion, the sample is atomized by flame, and the absorbance is measured at 324.8 nm. Within a certain concentration range, the absorbance of copper is proportional to its content, and quantification is performed by comparison with the standard series.

2 Ambito di applicazione

This method is applicable to the determination of copper in nylon.

3 Instruments and Reagents

3.1 Strumenti e attrezzature

3.1.1 Testing Instruments

- No, no, no. Nome Quantità Requisito tecnico Accessori
1 Spettrofotometro di assorbimento atomico di fiamma 1 set Copper Hollow-Cathode Lamp
2 compressore aria 1 set Pressione nominale di scarico: 0,3 MPa
3 Acetilene 1 cilindro Purity ≥99.99%