1 Resumo do Método
After sample digestion, the sample is atomized by flame, and the absorbance is measured at 324.8 nm. Within a certain concentration range, the absorbance of copper is proportional to its content, and quantification is performed by comparison with the standard series.
2. Âmbito de aplicação
This method is applicable to the determination of copper in nylon.
3 Instruments and Reagents
3.1 Instrumentos e equipamento
3.1.1 Testing Instruments
| Não. | Nome | Quantidade | Requisitos técnicos | Acessórios |
| 1 | Espetrofotometro de Absorção Atômica de Flame | 1 conjunto | — | Copper Hollow-Cathode Lamp |
| 2 | Compressor de ar | 1 conjunto | Pressão de escape estimada: 0,3 MPa | — |
| 3 | Acetileno | 1 cilindro | Purity ≥99.99% | — |