تحديد الرصاص في مياه الشرب (طيف الامتصاص الذري في فرن الجرافيت)

1 Method Summary

After appropriate pretreatment, test samples are atomized in graphite furnace, and absorbance is measured at the wavelength of 283.3 nm. Within a certain concentration range, the absorbance of lead is proportional to its mass concentration, and the content is quantified by comparison with standard calibration series.

2 Scope of Application

This method applies to the determination of lead in drinking water.

3 Reference Standard

GB/T 5750.6-2006 Standard Examination Methods for Drinking Water

4 Testing Instruments

· Graphite Furnace Atomic Absorption Spectrophotometer 1 set with Lead hollow cathode lamp