1 Yöntem Genel Bakış
The sample is heated and dissolved in a quartz beaker with mixed sulfuricphosphoric acid. Using an airacetylene flame, the absorbance of sodium oxide is measured at 589.0 nm on an atomic absorption spectrophotometer. Within a certain concentration range, the absorbance of sodium is proportional to its content, and quantification is carried out by comparison with the standard series.
2. Uygulama Kapsamı
This method is applicable to the determination of sodium oxide content in alumina powder.
3 Reference Basis
GB/T 6609.5-2004, Methods for chemical analysis and physical property testing of alumina — Determination of sodium oxide content, Method 2
4 Alet ve Reaktifler
4.1 Aletler ve Ekipmanlar
4.1.1 Testing Instruments
| - Hayır. - Hayır. | İsim | Miktar | Teknik Gereksinim | Aksesuarlar |
| 1 | Atomic Absorption Spectrophotometer (TAS990) | 1 set | — | Sodium Hollow-Cathode Lamp |
| 2 | Hava Kompresörü | 1 set | Nominal egzoz basıncı: 0.3 MPa | — |
| 3 | Asetilen | 1 silindir | Purity ≥99.99% | — |