Determination of Sodium Oxide in Alumina Powder (Flame Atomic Absorption Spectrometry)

1 Aperçu de la méthode

The sample is heated and dissolved in a quartz beaker with mixed sulfuricphosphoric acid. Using an airacetylene flame, the absorbance of sodium oxide is measured at 589.0 nm on an atomic absorption spectrophotometer. Within a certain concentration range, the absorbance of sodium is proportional to its content, and quantification is carried out by comparison with the standard series.

2. Champ d'application

This method is applicable to the determination of sodium oxide content in alumina powder.

3 Reference Basis

GB/T 6609.5-2004, Methods for chemical analysis and physical property testing of alumina — Determination of sodium oxide content, Method 2

4 Instruments et réactifs

4.1 Instruments et équipements

4.1.1 Testing Instruments

- Non, non. Nom Quantité Exigences techniques Accessoires
1 Atomic Absorption Spectrophotometer (TAS990) 1 ensemble Sodium Hollow-Cathode Lamp
2 Compresseur d'air 1 ensemble Pression nominale d'échappement: 0,3 MPa
3 Acetylène 1 cylindre Purity ≥99.99%