Determination of Iron in Zinc Alloys (Flame Atomic Absorption Spectrometry)

1 Обзор метода

The sample is digested with hydrochloric acid, hydrogen peroxide and nitric acid solution. In dilute hydrochloric acid medium, the absorbance of iron is measured by air‑acetylene flame at the wavelength of 248.3 nm using an atomic absorption spectrometer.

2. Область применения

This method is applicable for the determination of iron in zinc alloys.

3 Эталонный стандарт

GB/T 12689.5‑2004 Chemical analysis methods for zinc and zinc alloys — Determination of iron content — Sulfosalicylic acid spectrophotometry and flame atomic absorption spectrometry, Method 2: Flame atomic absorption spectrometry

4 Apparatus and Reagents

4.1 Apparatus and Equipment

4.1.1 Testing Instruments

Нет, нет. Имя Количество Технические требования Аксессуары
1 Спектрофотометр атомного поглощения пламени 1 набор Iron hollow‑cathode lamp
2 Воздушный компрессор 1 набор Rated discharge pressure: 0.3 MPa
3 ацетилен 1 цилиндр Purity ≥99.99%

контакт

Встретиться с профессионалами

Больше новостей

Оставить сообщение
Давайте получим разговор от нашей команды профессионалов

Документация о продуктах будет предоставлена ​​по электронной почте, если вы оставите свой адрес электронной почты.