Determination of Iron in Zinc Alloys (Flame Atomic Absorption Spectrometry)

1 Aperçu de la méthode

The sample is digested with hydrochloric acid, hydrogen peroxide and nitric acid solution. In dilute hydrochloric acid medium, the absorbance of iron is measured by air‑acetylene flame at the wavelength of 248.3 nm using an atomic absorption spectrometer.

2. Champ d'application

This method is applicable for the determination of iron in zinc alloys.

3 Norme de référence

GB/T 12689.5‑2004 Chemical analysis methods for zinc and zinc alloys — Determination of iron content — Sulfosalicylic acid spectrophotometry and flame atomic absorption spectrometry, Method 2: Flame atomic absorption spectrometry

4 Apparatus and Reagents

4.1 Apparatus and Equipment

4.1.1 Testing Instruments

- Non, non. Nom Quantité Exigences techniques Accessoires
1 Spectrophotomètre d'absorption atomique de flamme 1 ensemble Iron hollow‑cathode lamp
2 Compresseur d'air 1 ensemble Rated discharge pressure: 0.3 MPa
3 Acetylène 1 cylindre Purity ≥99.99%