1 Gambaran Umum Metode
The sample is digested with hydrochloric acid, hydrogen peroxide and nitric acid solution. In dilute hydrochloric acid medium, the absorbance of iron is measured by air‑acetylene flame at the wavelength of 248.3 nm using an atomic absorption spectrometer.
2. Ruang Lingkup Penerapan
This method is applicable for the determination of iron in zinc alloys.
3 Standar Referensi
GB/T 12689.5‑2004 Chemical analysis methods for zinc and zinc alloys — Determination of iron content — Sulfosalicylic acid spectrophotometry and flame atomic absorption spectrometry, Method 2: Flame atomic absorption spectrometry
4 Apparatus and Reagents
4.1 Apparatus and Equipment
4.1.1 Testing Instruments
| Tidak, tidak. | Nama | Kuantitas | Persyaratan Teknis | Aksesoris |
|---|---|---|---|---|
| 1 | Spektrofotometer penyerapan atom api | 1 set | — | Iron hollow‑cathode lamp |
| 2 | Kompresor udara | 1 set | Rated discharge pressure: 0.3 MPa | — |
| 3 | Acetilena | 1 silinder | Purity ≥99.99% | — |