1 Yöntem Genel Bakış
After microwave digestion, the sample solution is injected into the graphite furnace of the atomic absorption spectrophotometer. Following electrothermal atomization, it absorbs the resonance line at 283.3 nm. Within a certain concentration range, the absorbance is proportional to the lead content, and quantification is performed by comparison with the standard series.
2. Uygulama Kapsamı
This method is applicable to the determination of lead in graphite powder.
3 Reference Basis
GB/T 24533-2009
4 Alet ve Reaktifler
4.1 Aletler ve Ekipmanlar
4.1.1 Testing Instruments
| - Hayır. - Hayır. | İsim | Miktar | Teknik Gereksinim | Aksesuarlar |
|---|---|---|---|---|
| 1 | Grafit Fırın Atom Emme Spektrofotometri | 1 set | — | Lead Hollow-Cathode Lamp |
| 2 | Argon | 1 silindir | Purity ≥99.99% | — |
| 3 | Sirkülasyon Soğutma Su Sistemi | 1 set | Rated pump flow rate: 3.5 L/min | — |