アプリケーション

グラファイト粉末におけるググラファイトのグググラファイト炉原子吸収分光谱計 (Graphite Furnace Atomic Absorption Spectrometry)

1 方法概要

After microwave digestion, the sample solution is injected into the graphite furnace of the atomic absorption spectrophotometer. Following electrothermal atomization, it absorbs the resonance line at 283.3 nm. Within a certain concentration range, the absorbance is proportional to the lead content, and quantification is performed by comparison with the standard series.

2 適用範囲

This method is applicable to the determination of lead in graphite powder.

3 Reference Basis

GB/T 24533-2009

4.器具および試薬

4.1計器設備

4.1.1 Testing Instruments

いや 名前 技術要件 アクセサリー
1 グラファイト炉の原子吸収分光度計 1セット Lead Hollow-Cathode Lamp
2 Argon 1つのシリンダー Purity ≥99.99%
3 循環冷却水システム 1セット Rated pump flow rate: 3.5 L/min

接触