Tu400 vis
TU500 UV-vis
T6V Vis
T6U UV-vis
TU600 UV-vis
T7 UV-vis
T7S UV-vis
T7D UV-vis
TU700 UV-vis
T7DS UV-vis
T8DCS UV-vis
T9DCS UV-vis
T10DCS UV-vis
SOFTWARE UVWIN 6/GMP
Kit de qualificação UV/VIS IQ/OQ/PQ
Ftir8000
Ftir8100
A500
A3F
A3G
A3AFG
AA990F
AA990G
AA990AFG
PF7
FP912-2
FP912-3
FP912-4
FP912-5
AAS IQ/OQ/PQ
XD-2
XD-3
XD-6
M7 quadrupolo único GC-MS
G5 GC
GC1100 GC
L600 Líquido de alto desempenho
IPW-1000
I-Box
GBW-1
GWB-1-B
GWB-2
GWB-2-B
Sistema de digestão de microondas M40
Arruela de Labware D70E

Determinação do óxido de potássio em pó de alumínia

1 Resumo do Método
The test portion is heated and dissolved in a quartz beaker with mixed sulfuric‑phosphoric acid. After dilution with water, the absorbance of potassium is measured at a wavelength of 766.5 nm on an atomic absorption spectrophotometer using an air‑acetylene flame. Within a certain concentration range, the absorbance of potassium is proportional to its content, and quantification is performed by comparison with the standard series.

2. Âmbito de aplicação
This method is applicable to the determination of potassium oxide content in alumina powder.

3 Reference Basis
GB/T 6609.6‑2018, Methods for chemical analysis and physical property testing of alumina — Part 6: Determination of potassium oxide content

4 Instrumentos e Reagentes

4.1 Instrumentos e equipamento

4.1.1 Testing Instruments

Não. Nome Quantidade Requisitos técnicos Acessórios
1 Atomic Absorption Spectrophotometer (TAS‑990) 1 conjunto Potassium Hollow‑Cathode Lamp
2 Compressor de ar 1 conjunto Rated exhaust pressure: 0.3 MPa
3 Acetileno 1 cilindro Purity ≥99.99%

contato

Encontrar profissionais

Mais notícias

Deixe uma Mensagem
Vamos conseguir o Talkback da nossa equipe de profissionais

A documentação sobre os produtos será fornecida por e -mail de retorno se você deixar seu endereço de e -mail.