Determination of Lead in Olive Oil (Graphite Furnace Atomic Absorption Spectrometry)

1. Method Overview

After the sample is digested by microwave, it is injected into the graphite furnace of an atomic absorption spectrophotometer. After electrothermal atomization, the sample absorbs the resonance line at 283.3 nm. Within a certain concentration range, the absorbance value is proportional to the lead content, and quantification is performed by comparison with a standard series.

2. Instruments and Reagents

2.1 Instruments and Equipment

2.1.1 Testing Instruments

Serial No. 名前 Technical Requirements アクセサリー
1 Graphite Furnace Atomic Absorption Spectrophotometer 1 set Lead Hollow Cathode Lamp
2 Argon Gas 1 cylinder Purity ≥ 99.99%
3 Circulating Cooling Water System 1 set

2.1.2 Sample Pretreatment Equipment

Serial No. 名前 Technical Requirements アクセサリー
1 Microwave Digestion System 1 set Equipped with platinum resistor and infrared temperature measurement; pressure measurement accuracy: ±0.1℃
2 Temperature-Controlled Hot Plate 1 set
3 Electronic Balance 1 set Sensitivity: 0.1 mg
4 Ultrapure Water System 1 set
5 Volumetric Flask Several Volume: 100 mL
6 Test Tube Several Volume: 10 mL

2.2 Reagents

2.2.1 Raw Reagents

Serial No. 名前 Technical Requirements 備考
1 Diammonium Hydrogen Phosphate Analytical Reagent (AR)
2 Hydrochloric Acid Guaranteed Reagent (GR)
3 Nitric Acid MOS Grade
4 Palladium Nitrate Analytical Reagent (AR)

2.2.2 Prepared Reagents

Serial No. 名前 Preparation Method 備考
1 Nitric Acid Solution (1+99) Measure 10 mL of nitric acid, slowly add it to 990 mL of water, and mix well.
2 Nitric Acid Solution (5+95) Measure 5 mL of nitric acid, slowly add it to 95 mL of water, and mix well.
3 Diammonium Hydrogen Phosphate-Palladium Nitrate Solution Weigh 0.02 g of palladium nitrate, dissolve it with a small amount of nitric acid solution (1+9), then add 2 g of ammonium dihydrogen phosphate. After dissolution, dilute to 100 mL with nitric acid solution (5+95) and mix well.

2.3 Reference Standards

2.3.1 Stock Solution

Serial No. No. 名前 Technical Requirements 備考
1 GBW 08619 Lead Single-Element Standard Solution Concentration: 1000 μg/mL National Institute of Metrology, China

3. Operation Procedure

3.1 Sample Preparation

3.1.1 Preparation of Test Solution

Accurately weigh 0.2 g of the solid sample (precision to 0.001 g) into a microwave digestion vessel, add 5 mL of nitric acid, and digest the sample according to the operating procedures of the microwave digestion system. After cooling, take out the digestion vessel, and evaporate the acid on a hot plate at 140℃~160℃ until the volume is reduced to approximately 1 mL. After the digestion vessel cools down, transfer the digested solution to a 25 mL volumetric flask with deionized water. Wash the digestion vessel with a small amount of water for 2~3 times, combine the washing solutions into the volumetric flask, and dilute to the marked volume with water. Mix well for later use; this is the sample test solution. Prepare a reagent blank test at the same time.


Microwave Heating Program


Step Temperature/℃ Holding Time/min
1 100 3
2 120 3
3 140 3
4 160 3
5 180 30

3.1.2 Preparation of Standard Solutions

  1. Preparation of Lead Standard Intermediate Solution: For the lead standard solution (1.0 μg/mL), accurately pipette 0.1 mL of the lead standard stock solution (1000 μg/mL) into a 100 mL volumetric flask, dilute to the marked volume with (1+99) nitric acid solution, and shake well.
  2. Preparation of Lead Standard Series: Accurately pipette 0.0, 0.1, 0.2, 0.3, and 0.4 mL of the lead standard solution (1.0 μg/mL) into separate 10 mL volumetric flasks respectively. Dilute to the marked volume with (1+99) nitric acid solution to obtain the lead standard series solutions with concentrations of 0.0, 10.0, 20.0, 30.0, and 40.0 ng/mL.

3.2 Sample Testing

  1. Testing Conditions Reference testing conditions for the graphite furnace atomic absorption spectrophotometer: | Parameter | Specification (Lead) | | — | — | | Wavelength (nm) | 283.3 | | Spectral Bandwidth (nm) | 0.4 | | Element Lamp Current (mA) | 2 | | Background Correction Method | Deuterium Lamp |


Graphite Furnace Heating Program


Step 手順 Temperature (℃) Heating Time (s) Holding Time (s)
1 Drying 120 10 10
2 Ashing 650 10 15
3 Atomization 1900 0 2
4 Cleaning 2000 1 2


  1. Sample Testing In ascending order of mass concentration, inject 10 μL of the lead standard series solution and 5 μL of the diammonium hydrogen phosphate-palladium nitrate solution into the graphite furnace simultaneously. After atomization, measure the absorbance values. Plot a standard curve with mass concentration as the abscissa and absorbance value as the ordinate. Read the data, input the sample weight, and check and record the results.

3.3 Result Calculation

The lead content in the sample is calculated using the following formula:

どこ:
  • X = Lead content in the sample, unit: milligram per kilogram (mg/kg);
  • C1 = Mass concentration of lead in the sample test solution, unit: nanogram per milliliter (ng/mL);
  • C2= Mass concentration of lead in the blank solution, unit: nanogram per milliliter (ng/mL);
  • V = Volume of the sample after constant volume, unit: milliliter (mL);
  • m = Mass of the sample, unit: gram (g);
  • 1000 = Conversion factor.