검색
이 검색창을 닫습니다.

응용 프로그램

Determination of Vitamin B₂(Ultraviolet-Visible Spectrophotometry)

  1. Method Overview

This method is based on GB 12752-2010 National Food Safety Standard – Food Additive Vitamin B₂(Riboflavin). After dissolving vitamin B₂, it is diluted with sodium acetate solution, and a 1cm quartz cuvette is used for spectral scanning to determine the maximum absorption peak. The absorbance ratio at 375nm to that at 267nm should be 0.31~0.33, and the absorbance ratio at 222nm to that at 267nm should be 0.36~0.39.

  1. 2. Instruments and Reagents

2.1 Instruments and Equipment

2.1.1 Testing Instruments

Tu1901 UV Spectrophotometer(240nm~465nm)

1cm Quartz Cuvette

2.1.2 Pretreatment Equipment

Analytical Balance(0.001g precision)

2.2 Reagents

2.2.1 Reagents

Glacial acetic acid

Sodium acetate

2.2.2 Prepare Reagents

Sodium acetate solution(14g/L)

Weigh 1.4g of sodium acetate and dilute it to 100mL with deionized water.

  1. Operational Procedures

3.1 Sample Preparation

3.1.1 Preparation of Test Solution

Weigh 15mg of the powder sample, dissolve it in 5mL of acetic acid and 200mL of water, and dilute to 500mL.

3.2 Sample Testing

1)Testing Conditions

Spectral Bandwidth: 1nm

Scanning Range: 240nm~465nm

Speed: Medium

Interval: 1nm

2)Sample Testing

Scan the laboratory sample solution to identify maximum absorption at wavelengths of 444nm±1nm, 375nm±1nm, and 267nm±1nm, and measure the absorbance (A).

 