Determination of Iron in Graphite Powder (Flame Atomic Absorption Spectrometry)

1. Method Overview

After the sample is digested with aqua regia, the determination is carried out using an air-acetylene flame at an instrument wavelength of 248.3 nm.

2. Instruments and Reagents

2.1 Instruments and Equipment

2.1.1 Testing Instruments

Serial No. 이름 수량 Technical Requirements 부속품
1 Flame Atomic Absorption Spectrophotometer 1 set Iron Hollow Cathode Lamp
2 Acetylene Gas 1 cylinder Purity ≥ 99.99%
3 Air Compressor 1 set Rated discharge pressure: 0.3 MPa

2.1.2 Sample Pretreatment Equipment

Serial No. 이름 수량 Technical Requirements 부속품
1 Temperature-Controlled Hot Plate 1 set Rated temperature range: Room temperature ~ 300 ℃
2 Electronic Balance 1 set Sensitivity: 0.1 mg
3 Ultrapure Water System 1 set
4 Micropipette 1 each Ranges: 20 μL ~ 200 μL, 100 μL ~ 1000 μL
5 Volumetric Flask Several Volume: 100 mL
6 Test Tube Several Volume: 25 mL

2.2 Reagents

2.2.1 Raw Reagents

Serial No. 이름 Technical Requirements 발언
1 Nitric Acid MOS Grade
2 Hydrochloric Acid Guaranteed Reagent (GR)

2.2.2 Prepared Reagents

Serial No. 이름 Preparation Method 발언
1 (1+99) Nitric Acid Solution Measure 10 mL of nitric acid, slowly add it to 990 mL of deionized water, and mix well.

2.3 Reference Standards

2.3.1 Stock Solution

Serial No. No. 이름 Technical Requirements 발언
1 GBW08616 Iron Single-Element Standard Substance Solution Concentration: 1000 μg/mL Provided by National Institute of Metrology, China

3. Operational Procedures

3.1 Sample Preparation

3.1.1 Preparation of Test Solution

Weigh 0.25 g of the sample (accurate to 0.00001 g) into a digestion tank. Add 3 mL of concentrated nitric acid and 9 mL of concentrated hydrochloric acid, cover and tighten the digestion tank, then place it in a microwave digestion system for digestion. Prepare a blank sample using the same method. After cooling, filter the solution and dilute it to the marked volume in a 25 mL volumetric flask. The solution is ready for on-machine testing.
Microwave Heating Program
Step Temperature (℃) Holding Time (min)
1 100 3
2 120 3
3 140 3
4 160 3
5 180 30

3.1.2 Preparation of Standard Solutions

  1. Preparation of Iron Standard Series

    Accurately pipette 0.0, 0.05, 0.1, 0.2, and 0.3 mL of the iron standard solution (1000 μg/mL) into separate 100 mL volumetric flasks respectively. Add 5 mL of strontium chloride solution to each flask, then dilute to the marked volume with (1+99) nitric acid solution and mix well. The concentrations of this standard series are 0.0, 0.5, 1.0, 2.0, and 3.0 μg/mL.

3.2 Sample Testing

  1. Testing Conditions

    Reference Conditions for Flame Atomic Absorption Spectrophotometer

Parameter Specification
요소 Iron
Wavelength (nm) (To be specified based on instrument calibration)
Spectral Bandwidth (nm) (To be specified based on instrument calibration)
Element Lamp Current (mA) 3.0
Fuel Height (mm) 6
Fuel Flow Rate (mL/min) 1400
Air Compressor Pressure (MPa) 0.22
Background Correction Method None
  1. 샘플 테스트

    Adjust the instrument to its optimal working state, ignite the flame and adjust the fuel flow rate. Test the standard solutions and sample solutions in sequence, read the data, input the sample weight, then check and record the results.