- Method Overview
When iron ions in the sample are atomized, they absorb the resonance line (302.1nm) emitted by the iron hollow cathode lamp, and the absorption intensity is proportional to the iron content in the sample.
- Scope of Application
Applicable to the determination of iron in graphene leachate.
3. Detection Instruments
Graphite Furnace Atomic Absorption Spectrometer(Iron Hollow Cathode Lamp)
Cooling Circulating Water Device(Pump rated flow: 3.5L/min )
Argon Gas(1 bottle Purity ≥99.99% )
Balance