Determination Method for Lead in Surface Water (Graphite Furnace Atomic Absorption Spectrometry)

  1. Method Principle

After appropriate pretreatment, the sample is injected into a graphite tube atomizer. Lead ions are dissociated into atomic vapor at high temperatures within the graphite tube. The ground-state atoms of the target element absorb the resonance line emitted by a lead hollow cathode lamp, and the absorption intensity is proportional to the lead concentration within a certain range, enabling quantitative analysis of lead content.

  1. Scope

Applicable to the determination of lead in surface water.

  1. References

GB/T 5750.6-2006 Standard examination methods for surface water — Metal indices

  1. Testing Instruments
  • Graphite furnace atomic absorption spectrometer 1 set Lead hollow cathode lamp
  • Cooling circulating water system 1 set Rated flow rate of water pump: 3.5 L/min
  • Argon gas 1 cylinder Purity ≥ 99.99%