{"id":4479,"date":"2026-03-16T10:25:42","date_gmt":"2026-03-16T02:25:42","guid":{"rendered":"https:\/\/www.pgeneral.com\/?p=4479"},"modified":"2026-03-16T10:25:42","modified_gmt":"2026-03-16T02:25:42","slug":"determination-method-for-lead-in-surface-water-graphite-furnace-atomic-absorption-spectrometry-2","status":"publish","type":"post","link":"https:\/\/www.pgeneral.com\/fr\/applications\/environment\/determination-method-for-lead-in-surface-water-graphite-furnace-atomic-absorption-spectrometry-2","title":{"rendered":"M\u00e9thode de d\u00e9termination du plomb dans les eaux de surface (spectrom\u00e9trie d'absorption atomique en four graphite)"},"content":{"rendered":"<ol>\n<li>Principe de la m\u00e9thode<\/li>\n<\/ol>\n<p>Apr\u00e8s un pr\u00e9traitement appropri\u00e9, l'\u00e9chantillon est inject\u00e9 dans un atomiseur \u00e0 tube de graphite. Les ions de plomb sont dissoci\u00e9s en vapeur atomique \u00e0 haute temp\u00e9rature dans le tube de graphite. Les atomes \u00e0 l'\u00e9tat fondamental de l'\u00e9l\u00e9ment cible absorbent la ligne de r\u00e9sonance \u00e9mise par une lampe \u00e0 cathode creuse en plomb, et l'intensit\u00e9 de l'absorption est proportionnelle \u00e0 la concentration en plomb dans une certaine plage, ce qui permet une analyse quantitative de la teneur en plomb.<\/p>\n<ol start=\"2\">\n<li>Port\u00e9e<\/li>\n<\/ol>\n<p>Applicable \u00e0 la d\u00e9termination du plomb dans les eaux de surface.<\/p>\n<ol start=\"3\">\n<li>R\u00e9f\u00e9rences<\/li>\n<\/ol>\n<p>GB\/T 5750.6-2006 M\u00e9thodes d'examen normalis\u00e9es pour les eaux de surface \u2014 Indices m\u00e9talliques<\/p>\n<ol start=\"4\">\n<li>Instruments d'essai<\/li>\n<\/ol>\n<ul>\n<li>Spectrom\u00e8tre d'absorption atomique de four \u00e0 graphite 1 set Lampe cathodique creuse au plomb<\/li>\n<li>Syst\u00e8me d'eau circulante de refroidissement 1 ensemble D\u00e9bit nominal de la pompe \u00e0 eau: 3,5 L\/min<\/li>\n<li>Argon gaz 1 cylindre Puret\u00e9 \u2265 99,99%<\/li>\n<\/ul>","protected":false},"excerpt":{"rendered":"<p>Method Principle After appropriate pretreatment, the sample is injected into a graphite tube atomizer. Lead ions are dissociated into atomic vapor at high temperatures within the graphite tube. The ground-state atoms of the target element absorb the resonance line emitted by a lead hollow cathode lamp, and the absorption intensity is proportional to the lead [&hellip;]<\/p>\n","protected":false},"author":2,"featured_media":0,"comment_status":"closed","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"_acf_changed":false,"footnotes":""},"categories":[75],"tags":[],"class_list":["post-4479","post","type-post","status-publish","format-standard","hentry","category-environment"],"acf":[],"_links":{"self":[{"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/posts\/4479","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/users\/2"}],"replies":[{"embeddable":true,"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/comments?post=4479"}],"version-history":[{"count":0,"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/posts\/4479\/revisions"}],"wp:attachment":[{"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/media?parent=4479"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/categories?post=4479"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.pgeneral.com\/fr\/wp-json\/wp\/v2\/tags?post=4479"}],"curies":[{"name":"WP","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}