Determination of Vitamin B₂(Ultraviolet-Visible Spectrophotometry)

  1. Method Overview

This method is based on GB 12752-2010 National Food Safety Standard – Food Additive Vitamin B₂(Riboflavin). After dissolving vitamin B₂, it is diluted with sodium acetate solution, and a 1cm quartz cuvette is used for spectral scanning to determine the maximum absorption peak. The absorbance ratio at 375nm to that at 267nm should be 0.31~0.33, and the absorbance ratio at 222nm to that at 267nm should be 0.36~0.39.

  1. Scope

Vitamin B2

  1. References

GB 14752-2010 *National Food Safety Standard – Food Additive Vitamin B2 (Riboflavin)*

4. Testing Instruments

TU1901 UV Spectrophotometer(240nm~465nm)

1cm Quartz Cuvette