TU-1900/1901 UV-VIS Spectrophotometer    
 


--------------------------------------------------------------

 

The TU-1900 and TU-1901 is our company's highest performing research-grade UV-VIS spectrophotometers. Its optical system utilizes a double- beam dynamic feedback ratio photometric recording system, a unique full-transmission/full-reflection design, and a highly sensitive photomultiplier tube as a detector. All operations are controlled through the PC: automatic wavelength control, automatic switch light-source control, and automatic spectrum bandwidth control (TU1900 has set spectrum bandwidth). 

This instrument features easily interchangeable holder accessories and an open-type light source chamber design. The software is operated under the WINDOWS platform. The TU1900s UV-VIS spectrophotometers perform photometric measurements, spectrum measurements, and quantitative measurements. It is easily operated, features excellent performance, automatic 900nm-190nm wavelength scan, automatic peak-picking. Its convenient and easy data processing is very suitable for medical, chemical, environmental, agricultural, or biological analytical work.

--------------------------------------------------------------

Features

* Excellent performance. High-performance blazed holographic grating and the optimized CT-type monochromator minimizes stray light and widens the photometric range in the TU-1900s.

* Ideal baseline stability. Double-beam dynamic feedback ratio photometric recording system reasonably coupled with an advance design electric control system ensures high stability of the instrument baseline.

* High resolution. The unique optic design of full-transmission and full-reflection will guarantee a clean separation of the light source into a double beam while at the same time energy enhancement will maintain the light energy of both beams. This design reduces noise and guarantees high resolution.

* Accurate wavelength. The automatic wavelength system and the automatic light-source interchange systems ensure wavelength accuracy and high holistic performance of instrument.

* Easy accessories replacement. The detachable design of the sample chamber easily facilitates exchange of a wide range of optional accessories and ensures satisfactory analysis of various kinds.

Versatile UV-Win software

* User-friendly arrangement. The open design of the light source chamber, socket deuterium lamp and socket tungsten halogen lamp facilitates easy light source replacement, easy maintenance, and reduces operation error.

* Versatile application software. The UV-Win application software is made in the Windows format, offering an extended range of instrumental control and data processing.

* Quality components. Many key components are chosen from some of the world¨s best manufacturers; these include the deuterium lamp, PMT and holographic grating. Careful selection of these components ensures the instrument¨s dependability, accuracy, and maximizes lifespan.

--------------------------------------------------------------

Specifications

Optical System: Czerny-Turner(C-T) type monochromator, with high-resolution, diffraction holographic grating.
 Wavelength range   190nm ~ 900nm
 Spectral bandwidth  0.1 nm, 0.2 nm, 0.5nm, 1.0 nm, 2.0 nm and 5.0 nm (6 option, automatic changeover for TU-1901)
 2.0 nm (set bandwidth for TU-1901)
 Stray light    0.02%T (220nm NaI)
 Wavelength accuracy  +/-0.3 nm (automatic calibration)
 Wavelength  reproducibility  0.1nm
 Light source  interchange  Automatic light source change

 

 Photometric system: double-beam ratio recording system
 Photometric  method  Transmittance, absorbance, reflectance, energy, concentration.
 Photometric  range  -0.4~4.0Abs
 Photometric  accuracy  +/-0.3%T (0~100%T), +/-0.002Abs (0~0.5 Abs), +/-0.004Abs (0.5~1.0 Abs)
 Photometric  reproducibility  0.001Abs (0~0.5Abs), 0.002Abs (0.5~1.0 Abs)
 Baseline  flatness  +/-0.001Abs
 Baseline  stability  0.0004Abs/h (500nm, after preheating)

 

 Software: UVWin, a windows-operated software system 
 Photometric  measurement

measuring the transmittance or absorbance of 1-10 wavelengths, with the capacity to make mathematical calculations according to set formula.

 Spectrum scanning 

carrying out scanning of transmittance or absorbance spectral curves with the capacity to handle such data as peak value checkout,  derivative spectrum and spectrum calculation.

 Quantitative  determination

single, dual, triple wave-lengths and microanalysis, with up to 20 points 1-4 times curvilinear regression.

 Kinetic  Measurements

Time scanning of transmittance or absorbance on 1 to 10 specified wavelengths and handling data such as peak value checkout, spectral microanalysis and spectral calculation.

 Result delivery

Survey results from data or parameter file access, may be transmitted to other file processor or electronic tabulator for report generation.     

 

 Mainframe:    Well-designed high-performance spectrophotometer

 Light Source    

 Socket deuterium lamp and socket tungsten lamp  (with no adjustment needed after replacement).
 Detector           Photomultiplier tube

 Sample Chamber     

 With accessories like a four-cell sample holder and  an integrating sphere.
 Dimensions  587mm X 562mm X 260mm
 Weight  34kg

 

 Other Requirements:
 Power Source  voltage 110-230V, frequency 50-60Hz alternating  current

 

--------------------------------------------------------------
Note: The TU-1800 features all the advanced functions of the TU-1800S, but is simplified by removing the selectable bandwidth feature. The TU-1800 will remain at 2.0 nm set spectral bandwidth while the TU-1800S can analyze using four different bandwidths.