TU-1900/1901
UV-VIS Spectrophotometer
--------------------------------------------------------------

The
TU-1900 and TU-1901 is our company's highest performing
research-grade UV-VIS spectrophotometers. Its optical
system utilizes a double- beam dynamic feedback
ratio photometric recording system, a unique full-transmission/full-reflection
design, and a highly sensitive photomultiplier tube
as a detector. All operations are controlled through
the PC: automatic wavelength control, automatic
switch light-source control, and automatic
spectrum bandwidth control (TU1900 has set spectrum
bandwidth).
This
instrument features easily interchangeable holder
accessories and an open-type light source chamber
design. The software is operated under the WINDOWS
platform. The TU1900s UV-VIS spectrophotometers
perform photometric measurements, spectrum measurements,
and quantitative measurements. It is easily operated,
features excellent performance, automatic 900nm-190nm
wavelength scan, automatic peak-picking. Its convenient
and easy data processing is very suitable for medical,
chemical, environmental, agricultural, or biological
analytical work.
--------------------------------------------------------------
Features
*
Excellent performance. High-performance
blazed holographic grating and the optimized
CT-type monochromator minimizes stray light
and widens the photometric range in the TU-1900s.
*
Ideal baseline stability. Double-beam
dynamic feedback ratio photometric recording
system reasonably coupled with an advance
design electric control system ensures high
stability of the instrument baseline.
*
High resolution. The unique optic design
of full-transmission and full-reflection will
guarantee a clean separation of the light
source into a double beam while at the same
time energy enhancement will maintain the
light energy of both beams. This design reduces
noise and guarantees high resolution.
*
Accurate wavelength. The automatic
wavelength system and the automatic light-source
interchange systems ensure wavelength accuracy
and high holistic performance of instrument.
*
Easy accessories replacement. The detachable
design of the sample chamber easily facilitates
exchange of a wide range of optional accessories
and ensures satisfactory analysis of various
kinds.
|
Versatile
UV-Win software |
*
User-friendly arrangement. The open
design of the light source chamber, socket
deuterium lamp and socket tungsten halogen
lamp facilitates easy light source replacement,
easy maintenance, and reduces operation
error.
*
Versatile application software. The
UV-Win application software is made in the
Windows format, offering an extended range
of instrumental control and data processing.
*
Quality components. Many key components
are chosen from some of the world¨s best
manufacturers; these include the deuterium
lamp, PMT and holographic grating. Careful
selection of these components ensures the
instrument¨s dependability, accuracy, and
maximizes lifespan.
|
--------------------------------------------------------------
Specifications
| Optical
System: Czerny-Turner(C-T)
type monochromator, with high-resolution, diffraction
holographic grating. |
| Wavelength
range
|
190nm
~ 900nm |
| Spectral
bandwidth |
0.1
nm, 0.2 nm, 0.5nm, 1.0 nm, 2.0 nm and 5.0 nm
(6 option, automatic changeover for TU-1901) |
|
2.0
nm (set bandwidth for TU-1901) |
| Stray
light
|
0.02%T
(220nm NaI) |
| Wavelength
accuracy |
+/-0.3
nm (automatic calibration) |
| Wavelength
reproducibility |
0.1nm |
| Light
source interchange |
Automatic
light source change |
|
Photometric system: double-beam ratio recording
system |
| Photometric
method |
Transmittance, absorbance, reflectance, energy,
concentration. |
| Photometric
range |
-0.4~4.0Abs |
| Photometric
accuracy |
+/-0.3%T
(0~100%T), +/-0.002Abs
(0~0.5 Abs), +/-0.004Abs
(0.5~1.0 Abs) |
| Photometric
reproducibility |
0.001Abs
(0~0.5Abs), 0.002Abs (0.5~1.0 Abs) |
| Baseline
flatness |
+/-0.001Abs |
| Baseline
stability |
0.0004Abs/h
(500nm, after preheating) |
| Software:
UVWin, a windows-operated software system |
| Photometric
measurement
|
measuring
the transmittance or absorbance of 1-10
wavelengths, with the capacity to make mathematical
calculations according to set formula.
|
| Spectrum
scanning |
carrying
out scanning of transmittance
or absorbance spectral curves with the
capacity
to handle such data as peak value checkout,
derivative
spectrum and spectrum calculation.
|
| Quantitative
determination |
single, dual, triple wave-lengths and microanalysis,
with up to 20 points 1-4 times
curvilinear
regression.
|
| Kinetic
Measurements |
Time
scanning of transmittance
or absorbance on 1 to 10 specified
wavelengths and handling
data such as peak value checkout, spectral microanalysis
and spectral calculation.
|
| Result
delivery |
Survey
results from data or parameter file access,
may
be
transmitted to other file processor or electronic
tabulator for report
generation.
|
Mainframe: Well-designed high-performance spectrophotometer |
Light
Source |
Socket
deuterium lamp and socket tungsten lamp (with
no adjustment needed after replacement). |
| Detector |
Photomultiplier
tube |
Sample
Chamber |
With
accessories like a four-cell sample holder and
an integrating sphere. |
| Dimensions |
587mm
X 562mm X 260mm |
| Weight |
34kg |
| Other
Requirements: |
|
Power Source |
voltage
110-230V, frequency 50-60Hz alternating current |
--------------------------------------------------------------
Note:
The TU-1800 features all the advanced functions
of the TU-1800S, but is simplified by removing
the selectable bandwidth feature. The TU-1800
will remain at 2.0 nm set spectral bandwidth while
the TU-1800S can analyze using four different
bandwidths.
|