TU-1800PC/1800SPC
UV-VIS
Spectrophotometer
--------------------------------------------------------------
The
TU-1800PC/1800SPC is our company’s popular UV-VIS spectrophotometer.
Its optical system adopts a split-beam measurement optical
system, detector for two highly sensitive silicon diodes;
instrument operations are controlled through the PC,
using UV-WIN, a WINDOWS platform software. The instrument
features an 8-cell holder, accessible light-source chamber. TU-1800PC/1800PCS
UV-VIS spectrophotometer performs photometric measurements,
spectrum measurements, and quantitative measurements.
It
is easily operated, features excellent performance,
automatic 1100nm-200nm wavelength scan, abundant measurement
features, and data-processing applications. The practical
TU-1800 series is suitable for medical, chemical, environmental,
agricultural, or biological analytical work.
--------------------------------------------------------------
Features
*
Low Stray Light and Wide photometric range.
The high-performance blazed holographic grating
reduces stray light of the instrument and widens
the photometric range.
*
Enhanced baseline stability. The dual beam
monitoring ratio system enhances baseline stability.
*
Heightened resolution. The large-diameter
light path magnifies the instrument's energy,
reduces noise and enhances resolution.
*
Automatic sequential measurement. The automatic
eight-cell sample holder carries out automatic
measurement of eight samples in succession. It
also features a one-touch measurement of the dissolution
of six drug samples and a blank
*
User-friendly light source. The socket
deuterium lamp and socket tungsten lamp facilitates
easy light source replacement, easy maintenance,
and reduces operation error. |
|
*
Versatile application software. The
UV-Win application software is made in the
Windows format, offering an extended range
of instrumental control and data processing.
*
Quality components. Many key components
are chosen from some of the world’s best
manufacturers; these include the deuterium
lamp, PMT and holographic grating. Careful
selection of these components ensures the
instrument’s dependability, accuracy, and
maximizes lifespan.
|
Specifications
Optical System:
X-shaped monochromator, with
high-resolution, diffraction holographic grating.
|
Wavelength
range |
200nm ~ 1100nm |
Spectral
bandwidth |
0.5nm,
1.0 nm, 2.0 nm and 5.0 (for TU-1800PCS) |
| |
2.0
nm (for TU-1800PC) |
| Stray
light |
0.3
%T (220nm NaI) |
| Wavelength
accuracy |
+/-0.5
nm (automatic correction) |
Wavelength
reproducibility |
0.2nm |
Photometric system:
dual beam ratio monitoring system
|
Photometric
method |
Transmittance,
absorbance, energy and concentration. |
Photometric
range |
-0.3~3.0Abs
(0~200%T) |
| Photometric
accuracy |
+/-0.3%T
(0~100%T), +/-0.002
Abs (0~0.5 Abs), +/-0.004
Abs (0.5~1.0 Abs) |
| Photometric
reproducibility |
0.15
%T (0~100 %T), 0.001 Abs (0~0.5 Abs), 0.002 Abs (0.5~1.0
Abs) |
| Baseline
flatness |
+/-0.004
Abs (200nm~1100nm) |
| Baseline
stability |
0.002
Abs/h (500nm, after preheating) |
Software:
Windows platform UV-WIN
|
Photometric
measurement |
With
repetitive measurement function for measuring
the photometric values at 1~10 wavelengths and
make mathematical calculations based upon input
equations. Utilizes an eight-cell holder to quickly
make sequential calibrations and measure seven
sample's photometric values.
|
| Spectrum
scanning |
Record
sample's spectral curve and perform data processing.
Can scan and obtain spectrum characteristics of
within a wavelength of 200nm~1100nm. Can conduct
derivative, logarithmic, arithmetical calculations
on the spectral curve, or peak-picking and spectral
graph zoom operations.
|
| Photometric
accuracy |
Can
conduct single wavelength, double wavelength,
triple wavelength quantitative determination.
Can perform spectral curve fit on 1-24 standards.
Provided with measurement repetition function.
Utilize the eight-cell holder to quickly make
sequential calibrations and measure seven sample's
concentration.
|
| Kinetic
Measurements |
Time
scanning of transmittance
or absorbance on 1 to 10 specified wavelengths
and handling
data such as peak value checkout, spectral microanalysis
and spectral calculation.
|
| Result
delivery |
Save
data file and parameter file. Can use compiler
or data sheet share data.
|
Mainframe: A compact spectrophotometer
|
Light
Source |
Socket
deuterium lamp and socket tungsten lamp |
Detector |
A
pair of silicon photodiode |
| Sample
Chamber |
Eight-cell
sample holder |
| Dimensions |
540mm
X 400mm X 250mm |
| Weight |
25kg |
Other
Requirements:
|
Power
Source |
voltage
110-230V, frequency 50-60Hz alternating current |
|
|
|
|
|