TAS-986 Atomic Absorption Spectrophotometer  

--------------------------------------------------------------

 

The TAS-986 is our company's highest-grade fully automatic atomic absorption spectrophotometer. All operations are controlled through the PC, with the AA-Win software based on a WINDOWS platform. The design of flame burner and graphite furnace in one compact unit enables automatic interchanging of two analytical methods. 

There are two background correction methods, self-reversal background correction and deuterium lamp background correction, to ensure the most precise analytical results. A self aligning eight-lamp turret makes for user-friendly multi-elemental atomic measurements. The unique furnace heating design ensures uniform heating and guarantee measurement accuracy. All around automatic safety precautions ensure reliable operations.

--------------------------------------------------------------

Features

Fully automatic control

,The integration of flame burner and graphite furnace in one compact unit enables automatic interchanging between the flame mode and furnace mode using few simple keystrokes.

,Eight-Lamp turret for automatic positioning of each hollow cathode lamp.

,Automatic fuel gas(C2H2) flow rate adjustment to optimize combustion efficiency in element analysis.

,Automatic adjustment of burner height for optimized analysis conditions.

,Automatic Reflecting/Transmitting beamsplitter positioning (to lead the light of deuterium lamp into optical path) when the Deuterium Lamp Background Correction function is selected

,Automated wavelength scanning and peak-seeking

Automatic switchover of spectral bandwidth (5 choices).

Automatic negative voltage and lamp current adjustment, automatic AA and BG energy balance, automatic igniting and interlock protection in case of flameout.

Advanced Graphite furnace

,Uniformity of the graphite tube temperature reduces chemical interference and residue interference.

The PC-controlled heating program offers a multitude of heating method choices. The optical temperature control ensures rapid heating and extends the life of the graphite tube.

 

Versatile UV-Win software

  Safety Assurance

Flame Monitoring: Constant monitoring. When accidental flameout occurs, an alarm is given and the acetylene flow is stopped immediately

Pressure Monitoring: Constant air pressure detection. When abnormalities are detected, acetylene flow will be automatically cut off for safety fire extinguishing. The argon gas pressure monitor will disconnect the electricity to the graphite tube when an absence of the protective argon gas is detected.

Cooling water flow monitoring: The power supply to the furnace will only be activated for heating when the cooling water flow is determined enough to cool down the furnace body. 

Graphite Tube Fracture Protection: Power supply used for heating is automatically shut off and an alarm is given if a sudden graphite tube fracture is detected. 

 

--------------------------------------------------------------

Specifications

 Optical System: Czerny-Turner(C-T) type monochromator.
 Wavelength range   190nm ~ 900nm
 Spectral bandwidth  0.1 nm, 0.2 nm, 0.5nm, 1.0 nm, and 2.0  nm. (5 steps, with automatic changeover)
 Wavelength accuracy  +/-0.3 nm (automatic correction)
 Wavelength reproducibility  0.2 nm
 Resolution  over 0.3 nm
 Baseline Stability  0.005A/30 min

 

 Flame analysis
 Sensitivity  Cu 5µg/ml>0.850 Abs
 Burner Head  Titanium alloy burner
 Nebulizer  High-efficiency glass nebulizer
 Atomization Chamber  Corrosion-resistant material
 Position Adjustment  Automatic flame and furnace changeover
 Automatic flame burner optimum height setting
 Safety  Automatic ignition of mixing air-acetylene gas  with protection control.

 

 Graphite furnace analysis
 Characteristic Mass  (Cd)  0.8Pg
 Temperature Range  Ambient - 2650C
 Heating  voltage feedback control when drying and  ashing                            
 optical temperature control when atomizing        
 Heating program  up to 9 steps with choices of ramp, ladder and  full-power heating

 

 Background correction 
 Deuterium Lamp  Background  Correction  1.0 Abs
 Self-Reversal  Background  Correction 

 3.0 Abs

 

 Data processing

 Analytical method

 flame AA, graphite furnace, hydride, and flame  emission
 Determination  method   calibration curves (1~3 order), standard  addition method

 Repetition times 

 1~20 times with calculations of average value,  SD and RSD
 Result Printout  output of parameters, data, spectra and  calibration curves

 

 Mainframe
 Light Source  8 lamps, 2 lamps simultaneously lit (with one  preheat)
 Power Supply  220V/50Hz three-phase AC, 200W(mainframe),  5000W(graphite furnace)
 Size  mainframe 110cm x 50cm x 45cm, 75kg 
 graphite furnace power supply  system 50cm x 50cm  x 45cm, 50kg

 

--------------------------------------------------------------
Note: The TU-1800 features all the advanced functions of the TU-1800S, but is simplified by removing the selectable bandwidth feature. The TU-1800 will remain at 2.0 nm set spectral bandwidth while the TU-1800S can analyze using four different bandwidths.